The TY813 series monocrystalline silicon pressure / differential pressure transmitter is a high-accuracy intelligent instrument that measures pressure, differential pressure, liquid level, flow and density, converting the measured value into a 2-wire 4–20 mA electrical signal (HART, RS485, 0–5 V and 0–10 V outputs selectable). Spans run from 0.1 kPa up to 40 MPa with reference accuracy up to ±0.05%, making it suitable for automatic control in industrial production processes.
The transmitter is built around a monocrystalline (single-crystal) silicon pressure sensor chip manufactured with silicon micro-machining (MEMS) technology. The circuit uses a microprocessor core with advanced digital isolation in a modular layout, giving extremely high anti-interference performance and stability. A built-in temperature sensor performs temperature compensation, improving accuracy, reducing temperature drift and providing high reliability with strong self-diagnostics. The fully sealed, all-welded sensor uses a dual-diaphragm overload protection design with high overload capability.
Operators can remotely calibrate, configure and monitor the transmitter via HART communication. The series covers eight model types — TY813TG, TY813DP, TY813GP, TY813-Y, TY813-Z, TY813LT, TY813S1 and TY813S2 — and is widely used in petroleum, chemical, electric power, metallurgy, water treatment, brewing and many other industries.
Features
Industrial measuring instrument for differential pressure of liquids and gases; core technology is a monocrystalline silicon pressure sensor chip manufactured with silicon micro-machining (MEMS) technology
Fully sealed, all-welded monocrystalline silicon pressure sensor; eliminates the influence of temperature and static pressure; dual-diaphragm overload protection design with high overload capability
High-precision monocrystalline silicon sensing chip ensures accurate and stable measurement
Outstanding anti-interference and anti-vibration capability, suitable for complex and harsh industrial environments
High-quality materials and manufacturing for very high reliability; simple structure, easy installation — can be mounted directly on the process without additional supports or fixtures
Specially suited to differential-pressure-derived measurements such as flow and liquid level
00–20 m (fill 00 for direct-mount type). For S1/S2: H-side length 0–20 m (0 m = LT direct-mount, temperature ≤185 °C); L-side length 0–20 m (0 m = single-flange remote type, no flange on L side)
Fill fluid
A / B / C / D
A = DC200-1 silicone oil (-10~250 °C, 1.07 g/cm³); B = KN17 low-temperature oil (-90~120 °C, 0.914 g/cm³); C = KN3.2 high-temperature oil (0~380 °C, 1.1 g/cm³); D = coconut oil-M20 (-15~225 °C, 0.94 g/cm³)
1 = 1/4NPT external thread to M20×1.5 external thread, Ø14 welded pipe adapter; 2 = 1/2NPT external thread to M20×1.5 external thread, Ø14 welded pipe adapter; 3 = Ø8 mm pneumatic quick connector; 4 = Ø10 mm pneumatic quick connector; 5 = Ø4 mm ferrule fitting; 6 = Ø6 mm ferrule fitting; 7 = Ø8 mm ferrule fitting; 8 = T-type M20×1.5 external thread fitting
Valve manifold
F / S
F = 2-valve manifold; S = 3-valve manifold
Diaphragm seal flange
—
Add diaphragm seal flange, secondary oil filling (refer to diaphragm seal flange selection)
Flange dimension reference (flat flange, mm)
Flange
Standard
D1
D2
D3
D4
T1
Bolt holes n
Bolt hole Ød
DN25
ANSI150
110
79.4
66
34
12.7
4
16
DN25
PN10/16
115
85
66
34
18
4
14
DN25
PN25/40
115
85
66
34
18
4
14
DN25
PN64
140
100
66
34
24
4
18
DN50
ANSI150
150
120.7
100
61
19.5
4
18
DN50
PN10/16
165
125
100
61
18
4
18
DN50
PN25/40
165
125
100
61
20
4
18
DN50
PN64
180
135
100
61
26
4
22
DN80
ANSI150
190
152.4
130
89
24.3
4
18
DN80
PN10/16
200
160
130
89
20
8
18
DN80
PN25/40
200
160
130
89
24
8
18
DN80
PN64
215
170
130
89
28
8
22
DN100
ANSI150
230
190.5
155
115
24.3
8
18
DN100
PN10/16
220
180
155
115
20
8
18
DN100
PN25/40
235
190
155
115
24
8
22
DN100
PN64
250
200
155
115
30
8
26
Matching dimension tables for extended (protruding) flanges (DN50/DN80/DN100) are available — contact us for the full drawing set.
Applications
Petroleum and petrochemical processing
Chemical industry
Electric power plants
Metallurgy
Water and wastewater treatment
Brewing and food/sanitary processes (LT type)
Flow measurement via differential pressure (orifice/venturi)
Liquid level measurement in vessels, including viscous and easily crystallizing media (flange / remote-seal types)
High-temperature media measurement (capillary remote-seal types, up to 420 °C)
FAQ
Which output signals does the TY813 support?
The TY813 is a 2-wire 4–20 mA transmitter; ordering options include standard 4–20 mA (code E), 4–20 mA + HART (code S), RS485 (code R) and 0–10 V voltage output (code V). HART allows remote calibration, configuration and monitoring.
Can the TY813 be installed in hazardous areas?
Yes. Besides the non-Ex version, the TY813 is available as intrinsically safe Ex ia IIC T6 Gb or flameproof Ex d IIC T6 Gb, with IP66/IP67 ingress protection and cast aluminum, 304 or 316 stainless steel housings.
How do I select the right TY813 model for my application?
Contact us with your medium, measuring range, temperature and process connection; we respond within 24 hours with a complete ordering code, including flange/remote-seal options, fill fluid and mounting accessories.